UES24LCP6-SPA
The UES24LCP6-SPA is a medical wall-plug adapter for equipment needing a dedicated DC supply. Choose the output and device connection for your design, with interchangeable AC plugs for different markets.

Choose the output voltage first, then check the current available at that voltage. The rows below describe separate output configurations; this adapter does not switch between them during use.
| Model Number | Power (W) | Input Voltage Range (VAC) | Output Voltage (VDC) | Output Current (A) | Operating Temperature Range (°C) | Safety Standards |
| UES24LCP6-XXXYYYSPA | 17.7 | 80–264 | 5.0–5.9 | 0.01–3.00 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 22.5 | 80–264 | 6.0–7.5 | 0.01–3.00 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 21 | 80–264 | 7.6–8.0 | 0.01–2.62 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 21 | 80–264 | 8.1–8.9 | 0.01–2.36 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 9.0–10.0 | 0.01–2.40 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 10.1–11.0 | 0.01–2.18 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 11.1–12.0 | 0.01–2.00 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 12.1–13.0 | 0.01–1.84 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 13.1–14.0 | 0.01–1.71 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 14.1–15.0 | 0.01–1.60 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 19.0–20.0 | 0.01–1.20 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 20.1–21.0 | 0.01–1.14 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 21.1–22.0 | 0.01–1.09 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 22.1–23.0 | 0.01–1.04 | 0–+40 | CB, UL, TÜV Mark |
| UES24LCP6-XXXYYYSPA | 24 | 80–264 | 23.1–24.0 | 0.01–1.00 | 0–+40 | CB, UL, TÜV Mark |

| Input Characteristic | Specification |
| Input Voltage Range | 80–264VAC |
| Frequency Range | 47–63Hz |
| Input Current | 0.7A at 100VAC |
| Inrush Current | 70A max. at 240VAC, cold start |
| Touch Leakage Current | ≤100μA at 264VAC |
| Characteristic | Specification |
| Operating Temperature | 0°C to +40°C |
| Storage Temperature | −20°C to +60°C |
| Operating Humidity | 10–90% RH, non-condensing |
| Storage Humidity | 5–90% RH |
| Operating Altitude | Up to 5,000m |
| Enclosure Rating | IP42 |
| Construction | Class II |
| Energy Efficiency | Level VI |
| Standby Power | ≤0.075W |
| MTBF | >100,000 hours, Telcordia SR-332 at 25°C |
| Protection Function | Specification |
| Overload | 110–200% of rated output power; automatic recovery |
| Overvoltage | 110–150% of rated output voltage; input reset required |
| Short Circuit | Trip and restart, hiccup mode |
IEC60601-1
View Certificate| EMC Item | Medical Specification |
| Conducted Emissions | IEC/EN 60601-1-2, CISPR 11 |
| Radiated Emissions | IEC/EN 60601-1-2, CISPR 11 |
| Harmonic Currents | EN 61000-3-2, Class A |
| Voltage Flicker | EN 61000-3-3 |
| Immunity Reference | IEC/EN 60601-1-2 |
| ESD | IEC 61000-4-2; ±15kV air, ±8kV contact |
| Radiated Immunity | IEC 61000-4-3; 10V/m, 3V/m, 80MHz–2.7GHz |
| EFT/Burst | IEC 61000-4-4; ±2kV on AC port, ±1kV on signal ports |
| Surge | IEC 61000-4-5; ±1kV line-to-line, differential mode |
| Conducted Immunity | IEC 61000-4-6; 3Vrms, 6Vrms, 0.15–80MHz |
| Magnetic Field | IEC 61000-4-8; 30A/m |
| Dips & Interruptions | IEC 61000-4-11; 0%, 70%, 0% of UT |
The UES24LCP6-SPA is a candidate for small medical devices that use an external DC supply. These applications include motor loads, monitoring electronics and lighting controllers, so they need different checks during selection. The chosen output must cover the complete device load, including charging where applicable; the application name alone does not establish compatibility.




If the required output or installation does not fit this adapter, compare the models below before choosing a sample. A different series may offer the current capacity, temperature range or physical format your equipment needs. For a replacement project, compare the complete specification alongside the connection details.
A product table helps narrow the choice, but some decisions need a closer look. These articles explain current capacity, connection details and medical electrical safety—the points worth settling before an adapter becomes part of the equipment design.
Considering the 12V/2A version for an existing design? This article explains the difference between current drawn by the device and current available from the adapter, with examples of how that affects replacement and load selection.
Use this overview when specifying the connection at either end of the adapter. It covers regional AC plugs, common DC connector sizes and polarity, helping avoid a sample that has the right output but the wrong physical connection.
For readers reviewing medical approval requirements, this article introduces the safety issues addressed by IEC 60601-1, including insulation, leakage current and fault conditions. It provides background for discussions between purchasing, design and certification teams.
Our engineering team provides comprehensive support for this power supply series. Whether you need standard models, custom DC output specifications, or specific medical compliance assistance, we ensure your requirements are met with precision and speed.
✔ 2 MOPP & IEC 60601-1 Certification Support
✔ Custom DC Connectors & Cable Options
✔ Free Engineering Samples for Evaluation
✔ Detailed Technical Response within 24h