- UES15D-SPA | 5V-24V 15W Medical Open Frame Power Supply
UES15D-SPA | 5V-24V 15W Medical Open Frame Power Supply
UES15D-SPA Series
The series combines a 15W power rating with 2 MOPP input-to-output isolation and touch leakage current ≤100μA.
- Seven fixed outputs: 5V, 6V, 9V, 12V, 15V, 18V and 24V
- 54.0 × 36.8 × 20.3mm body, weighing 48g
- PCB-mount pins for installation inside the equipment
- 2 MOPP input-to-output isolation and 4,000VAC dielectric withstand voltage
- Touch leakage current ≤100μA at 264VAC
- Standby power ≤0.25W
- 85–265VAC input, with an operating temperature range of −25°C to +65°C

Models & Ratings
| Model Number | Power(W) | Input Voltage Range(VAC) | Output Voltage(VDC) | Output Current(A) | Operating Temperature Range(°C) | Safety Standards |
| UES15D-XXXYYYSPA | 15 | 85–265 | 5 | 0.01–3.00 | −25 to +65 | CB, UL, CE, TÜV Mark |
| UES15D-XXXYYYSPA | 15 | 85–265 | 6 | 0.01–2.50 | −25 to +65 | CB, UL, CE, TÜV Mark |
| UES15D-XXXYYYSPA | 15 | 85–265 | 9 | 0.01–1.67 | −25 to +65 | CB, UL, CE, TÜV Mark |
| UES15D-XXXYYYSPA | 15 | 85–265 | 12 | 0.01–1.25 | −25 to +65 | CB, UL, CE, TÜV Mark |
| UES15D-XXXYYYSPA | 15 | 85–265 | 15 | 0.01–1.00 | −25 to +65 | CB, UL, CE, TÜV Mark |
| UES15D-XXXYYYSPA | 15 | 85–265 | 18 | 0.01–0.83 | −25 to +65 | CB, UL, CE, TÜV Mark |
| UES15D-XXXYYYSPA | 15 | 85–265 | 24 | 0.01–0.62 | −25 to +65 | CB, UL, CE, TÜV Mark |
Mechanical Details

Input Specifications
| Input Characteristic | Specification |
| Input Voltage Range | 85–265VAC |
| Frequency Range | 47–63Hz |
| Input Current | 0.5A at 100VAC |
| Inrush Current | 100A max at 240VAC, cold start |
| Standby Power | ≤0.25W |
If your product has its own mains switch, fuse or input filter, include those components in the input design review.
Environmental & General
| Operating Temperature | −25°C to +65°C |
| Storage Temperature | −25°C to +85°C |
| Operating Humidity | 10–90% RH, non-condensing |
| Storage Humidity | 10–90% RH |
| Operating Altitude | Up to 5,000m |
| MTBF | >100,000 hours, Telcordia SR-332 at 25°C |
The listed >100,000-hour MTBF is calculated using Telcordia SR-332 at 25°C. It is a reference for reliability assessment—not a promise that an individual supply will run for that many hours.
Protection
| Overload | 110–180% of rated output power; automatic recovery |
| Overvoltage | Catalog specifies 150% max of output voltage; input reset required |
| Short Circuit | Output trips and attempts to restart in hiccup mode |
Safety Approvals
IEC60601-1
View CertificateEN60601-1
View CertificateEMC
| Item | Medical | ITE |
| Conducted Emissions | IEC/EN 60601-1-2; CISPR 11 | EN 55032; CISPR 32 |
| Radiated Emissions | IEC/EN 60601-1-2; CISPR 11 | EN 55032; CISPR 32 |
| Harmonic Currents | EN 61000-3-2, Class A | EN 61000-3-2, Class A |
| Voltage Flicker | EN 61000-3-3 | EN 61000-3-3 |
| Immunity | IEC/EN 60601-1-2 | EN 55035; CISPR 35 |
Common Applications
Frequently Asked Questions
Is the UES15D-SPA suitable for my device?
First, what output voltage and current does your device need? Do you need any certifications, and which ones are required for your project?
If you already have a power supply, you can send us a photo of its label along with your device’s power requirements. That gives us a starting point for checking the output and certification requirements. Since the UES15D-SPA mounts inside the equipment, we’ll also need to know how much installation space you have.
With those details, we can check whether this model fits or whether another model would be a better choice.
I’m looking for a replacement. Can I keep the same PCB?
Possibly. Let’s compare the drawings before you make any changes.
The UES15D-SPA has four pins, with different spacing on the AC and DC sides. Matching the output is the first step; the pin positions and assignments also need to match your board. A photo of the existing supply is useful, but a drawing will give us a much clearer answer.
Would this be a good choice for a small pump?
With a pump, we’d look at the starting current as well as the current once it is running. The two figures can be quite different.
What pump are you using? Please send us its model number or a photo of the label. We may also need the starting-current figure, since that is not always shown on the label.
Can we order different outputs for different products?
Yes. Tell us the voltage and current needed for each product, and we can check which versions match.
For example, one product might need 6V at 2A, while another needs 9V at 1.5A. Both fall within the corresponding versions’ listed current ratings. They would be two different units, though—you cannot change one unit between those outputs.
How much room does it need inside our device?
The body is 54.0 × 36.8 × 20.3mm. Do you have a layout drawing showing where you want to put it? A sketch with dimensions is also useful.
We’ll look at the pin locations and the space around the supply as well as the body size. Please show any nearby components too, particularly anything that gets warm during use.
Can our device work and charge its battery at the same time?
How much current does it use while charging? That is the figure we need alongside the device’s operating current.
If your existing test only covers the device running with a fully charged battery, it may not show the highest load. Try to include the condition where the device is working and the battery is charging. The battery also needs its own charging circuit; that function is not built into the UES15D-SPA.
Related Models
Related Resources
Still checking whether this supply fits your project? These articles explain the medical isolation terms, installation details and safety documents you may need to review before making a choice.
What does 2 x MOPP mean for a medical power supply?
You’ll find 2 MOPP in the UES15D-SPA specifications. What does it mean for your device? This article looks at patient protection and the insulation requirements behind the term, giving you some background before reviewing the supply with your compliance team.
How to Install Open Frame AC-DC Power Supplies
Finding room for the power supply is one part of the installation. Wiring, cooling and electrical clearances need space too. This guide walks through those considerations and provides a checklist to review before closing the equipment enclosure.
12V 1A vs 12V 2A Adapter: Which One Should You Choose?
Two supplies can have the same voltage and very different current ratings. If you’re comparing the UES15D-SPA’s 12V version with another supply, this article explains what those current figures mean, why a higher rating does not force extra current into the device, and where starting demand matters.