Quankang | Professional Medical Power Supply Manufacturer

UES240E-SPCZ / UES240F-SPCZ | 5V–48V 240W Medical Desktop Power Adapter

UES240E-SPCZ / UES240F-SPCZ

Medical USB-C PD power up to 240W, with higher-voltage profiles for equipment designed to use 36V or 48V inputs.

  • 180W at 36V; 240W at 48V
  • USB PD 3.2, AVS and PPS support
  • Type-C socket or fixed output lead
  • 2 MOPP input-to-output isolation
  • Medical and home healthcare approval references
UES240E-SPCZ UES240F-SPCZ 5V-48V 240W Medical Desktop Power Supply/GaN PD Fast Charger

Models & Ratings

Start with the input your equipment needs. A 48V design can use the listed 240W profile; a 36V design has up to 180W available. At 20V, the limit is 100W.

E and F share these output ratings. The choice between them concerns the output connection, so you can settle the electrical requirement first and then choose the cable arrangement. The table below shows the fixed profiles. AVS and PPS requirements need checking separately because their ranges are not given in the supplied specification.

Model NumberPower(W)Input Voltage Range(VAC)Output Voltage(VDC)Output Current(A)Operating Temperature Range(°C)Safety Standards
UES240E-SPCZ / UES240F-SPCZ1590–264 / 80–264*50.01–3.00−10 to +40CB, UL, TÜV-SÜD/Mark
UES240E-SPCZ / UES240F-SPCZ2790–264 / 80–264*90.01–3.00−10 to +40CB, UL, TÜV-SÜD/Mark
UES240E-SPCZ / UES240F-SPCZ3690–264 / 80–264*120.01–3.00−10 to +40CB, UL, TÜV-SÜD/Mark
UES240E-SPCZ / UES240F-SPCZ4590–264 / 80–264*150.01–3.00−10 to +40CB, UL, TÜV-SÜD/Mark
UES240E-SPCZ / UES240F-SPCZ10090–264 / 80–264*200.01–5.00−10 to +40CB, UL, TÜV-SÜD/Mark
UES240E-SPCZ / UES240F-SPCZ14090–264 / 80–264*280.01–5.00−10 to +40CB, UL, TÜV-SÜD/Mark
UES240E-SPCZ / UES240F-SPCZ18090–264 / 80–264*360.01–5.00−10 to +40CB, UL, TÜV-SÜD/Mark
UES240E-SPCZ / UES240F-SPCZ24090–264 / 80–264*480.01–5.00−10 to +40CB, UL, TÜV-SÜD/Mark
Input range: Class I, 90–264VAC; Class II, 80–264VAC.

All profiles have ±5% voltage tolerance, ±2% line regulation, ±5% load regulation and a start-up delay of ≤3 seconds.

Maximum ripple and noise is 250mVp-p at 5–15V, 300mVp-p at 20V, 350mVp-p at 28V and 36V, and 400mVp-p at 48V.

Mechanical Details

If you want the output cable supplied separately, choose E. If you want it attached to the adapter, choose F. That is the main difference between the two versions.

Both bodies measure 150 × 73 × 32.5mm. When checking the space, include the plugs and cable bends too. For F, tell us the lead length you need—the drawing does not specify one.
UES240E SPCZ UES240F SPCZ Mechanical Details

Input Specifications

There is an input-range difference to keep in mind when choosing the mains inlet. C6 and C14 versions are specified for 90–264VAC. C8 and C18 versions are specified for 80–264VAC.
 
So, if your equipment specification calls for operation down to 80VAC, the inlet choice needs attention before you order. Use the configuration code shown below to identify the version required.
Input CharacteristicSpecification
Input voltage — Class I90–264VAC
Input voltage — Class II80–264VAC
Frequency range47–63Hz
Input current4.0A maximum at 80VAC
Standby power≤0.15W
AC Inlet Options
C8 / Polarized C8 — Class II; replace Z with 1
C6 — Class I; replace Z with 2
C14 — Class I; replace Z with 3
C18 — Class II; replace Z with 4

Environmental & General

Will the adapter be left on a desk, or fitted into your equipment? That makes a difference to the temperature it will see. For this series, the operating limit is +40°C around the adapter. If it sits beside a warm enclosure, that space can be hotter than the room.

For transport and storage, the specified range is −20°C to +80°C. The adapter must be switched off under those storage conditions; the higher temperature is not an operating allowance.
CharacteristicSpecification
Operating temperature−10°C to +40°C
Storage temperature−20°C to +80°C
Operating humidity10–90% RH, non-condensing
Storage humidity5–90% RH
Operating altitudeUp to 5,000m
Enclosure ratingIP42
Weight650g ±15g
Energy efficiencyLevel VII, as listed in the specification
MTBF>100,000 hours, Telcordia SR-332 at 25°C
Output ProfileAverage Efficiency
5V82.00%
9V87.30%
12V88.30%
15V88.90%
20V89.50%
28V90.00%
36V91.00%
48V91.50%

Protection

The output can come back on its own after an overload clears. A short circuit produces a repeated restart response while the fault remains. Overvoltage protection works differently: the AC input has to be disconnected and reapplied to reset it.

If your equipment keeps switching on and off, leave it disconnected while the cause is investigated. Restarting is a protection response, so repeatedly cycling the equipment is not something to accept during sample testing. Note what the equipment was doing when it stopped—starting a motor, charging a battery or changing operating mode—so we have a useful starting point for the investigation.
Protection FunctionSpecificationRecovery
Overload105–140% of rated output powerAutomatic recovery
Overvoltage110–130% of rated output voltageAC input reset required
Short circuitTrip and restart, hiccup modeRestart attempts while the fault remains

Safety Approvals

For a home healthcare project, include the IEC 60601-1-11 requirement when asking for documents. This specification lists it alongside the general medical safety standards.

The approval references below cover CB, UL and TÜV-SÜD/Mark. We’ll need the selected E/F version and mains inlet to check the relevant document coverage. Keep those details with the approved sample so purchasing has a clear configuration to order.
CB

IEC60601-1

View Certificate
TÜV

EN60601-1

View Certificate
UL

60601-1

View Certificate

EMC

For E, choose the separate output cable before completing the equipment’s EMC work. Keep that cable on the production parts list once it has been approved with the equipment.

For F, test with the fixed lead arranged as it will be used. A photo of the wiring is worth keeping with the results—it makes a later comparison much easier if the equipment layout changes. The stated medical test references and levels follow below.
ItemMedical Specification
Conducted emissionsIEC/EN 60601-1-2, CISPR 11
Radiated emissionsIEC/EN 60601-1-2, CISPR 11
Harmonic currentsEN 61000-3-2, Class A
Voltage flickerEN 61000-3-3
ImmunityIEC/EN 60601-1-2
Immunity TestStandardSpecified Level
ESDIEC 61000-4-2±15kV air, ±8kV contact
Radiated immunityIEC 61000-4-33V/m
EFT/BurstIEC 61000-4-4±2kV AC port, ±1kV signal ports
SurgeIEC 61000-4-5±2kV line-to-neutral; ±4kV line/neutral-to-ground
Conducted immunityIEC 61000-4-63Vrms, 6Vrms; 0.15–80MHz
Magnetic fieldIEC 61000-4-830A/m
Dips and interruptionsIEC 61000-4-110%, 70%, 0% of UT
Refer to the applicable report for frequency ranges, durations and test conditions.

Common Applications

These models can be evaluated for medical equipment designed to accept compatible USB-C PD power. Select from the device’s input requirements, including the power needed while operating and charging at the same time.

The categories below are application directions. They do not mean every device in the category accepts PD or requires 240W. Existing equipment with a conventional DC input needs its connection and power-input design reviewed before replacement.
Portable Ultrasound Diagnostic Equipment power supply
Portable Ultrasound Equipment
Medical grade display power supply
Medical Displays
Intraoral Scanners
Intraoral Scanners
Patient vital signs monitoring power supply
Patient Monitoring Equipment

Frequently Asked Questions

We’re moving our medical equipment to USB-C. What do you need to check before recommending this model?
What voltage and current does the equipment need at its power input? If you’re replacing an existing adapter, its label is a useful place to start.

We also need to know how your board will request power over USB-C. A matching connector alone won’t establish compatibility. Your hardware engineer may already have the PD input specification; include that with the medical approvals your project requires. We can then review whether this series is a suitable sample candidate.
We wouldn’t recommend changing to 240W on that symptom alone. First, we need to establish whether the reset happens during voltage negotiation or when the equipment starts drawing more current.

A voltage trace taken at the equipment input would help here. Include the requested PD profile and the cable used for the test. This series’ specification does not give a voltage-transition interruption time, so we would need to investigate the behaviour with your board before recommending it as a replacement.
Yes, all from one output—48V at 5A. There aren’t several ports dividing up the power.

Where people can get caught out is using it with a lower-voltage device. A device requesting 20V gets up to 100W from this adapter. It won’t get 240W just because that’s the number on the adapter.
For a 48V USB PD EPR input, yes, it’s a suitable option to try. A 200W load draws about 4.17A at 48V, and this output is rated for 5A.

If your equipment needs 36V, we’d choose another supply. This one gives 180W at 36V, which leaves you short. And if you’re replacing an ordinary 48V DC adapter, this won’t be a straight swap: the equipment needs a PD receiving circuit to request the output.
You’ll need a 240W EPR cable to get the full output from the E version. Keep the 100W cable for an application it’s rated for.

It’s easy to buy the wrong replacement here. Both cables can say “5A”, but the old 100W rating is based on 20V. For this adapter’s full output, the cable has to support 48V as well. Look for the 240W EPR rating rather than just the current marking.

With F, the lead is already attached.
I’d choose E if being able to change the output cable matters. A damaged removable cable can be replaced on its own, and you can pack the cable separately from the adapter.
 
F suits a kit where you want the adapter and lead kept together. Whoever unpacks it has the output lead there already.

You get the same listed power options with either version. There’s no power advantage to choosing F over E; it comes down to which arrangement works better for the people using and servicing your equipment.

Related Models

Start with the output profile when comparing alternatives. Equipment operating within a lower-power model’s ratings may not gain additional usable power from a 240W supply. A design requiring 36V or 48V needs those profiles checked explicitly.

The models below provide other ratings and connection arrangements to consider. Compare them with the equipment input and package requirements before choosing samples, especially when replacing a supply in an existing product.
UES140AZ-SPC/UES140BZ-SPC 5V-28V 140W Medical Desktop Power Supply /GaN PD Fast Charger

UES140AZ-SPC / UES140BZ-SPC | 5V–28V 140W Medical Desktop Power Adapter

120w medical power supply

UED120-SPA | 19V-24V 120W Medical Desktop Adapter

UES120DZ-SPA 11V-54V 120W Medical Desktop Power Adapter

UES120DZ-SPA | 11V-54V 120W Medical Desktop Adapter

UES100DZ-SPA 12V-24V 100W Medical Desktop Power Supply

UES100DZ-SPA | 12V-24V 100W Medical Desktop Adapter

Related Resources

These articles provide background for decisions beyond the ratings table: introducing USB-C power into medical equipment, allowing for installation conditions, and understanding patient-protection isolation.

Use them to prepare the discussion with your hardware and certification teams. Final selection should follow this model’s own specification and applicable reports, rather than parameters quoted for another product.

Medical grade USB-C power supply

An introduction to USB-C power in medical equipment, covering connection formats, power delivery and medical-use considerations. Useful when defining an external power arrangement and deciding what compatibility information to obtain from the equipment designer.

What does 2 x MOPP mean for a medical power supply?

Both models list 2 MOPP input-to-output isolation. Read what the term means and how it relates to patient protection before reviewing the isolation requirements and supporting documents with your medical equipment certification team.

Power Supply Derating: How to Calculate Output Limits for Temperature, Altitude, and Cooling

Review how operating conditions affect available power and why the installation belongs in the selection process. The article explains general derating principles; use the selected adapter’s documentation and test results to establish its actual limits.

Request Technical Support & Pricing

Our engineering team provides comprehensive support for this power supply series. Whether you need standard models, custom DC output specifications, or specific medical compliance assistance, we ensure your requirements are met with precision and speed.

✔ 2 MOPP & IEC 60601-1 Certification Support

✔ Custom DC Connectors & Cable Options

✔ Free Engineering Samples for Evaluation

✔ Detailed Technical Response within 24h

Request a Free Quote

Send us a message if you have any questions or request a quote. We will be back to you ASAP!