- UES140AZ-SPC / UES140BZ-SPC | 5V–28V 140W Medical Desktop Power Adapter
UES140AZ-SPC / UES140BZ-SPC | 5V–28V 140W Medical Desktop Power Adapter
UES140AZ-SPC / UES140BZ-SPC
A GaN desktop power supply for medical equipment using USB Power Delivery, available with a Type-C output port or a fixed output cable.
- USB PD 3.1 support for power requirements beyond 100W
- Type-C port or fixed-cable output
- Medical isolation with low touch leakage current
- Class I and Class II AC inlet configurations
- Low standby consumption for mains-connected equipment

Models & Ratings
Both models support the profiles below. The connected equipment requests its operating voltage through PD negotiation. Full 140W output is available at 28V/5A.
| Model Number | Power(W) | Input Voltage Range(VAC) | Output Voltage(VDC) | Output Current(A) | Operating Temperature Range(°C) | Safety Standards |
| UES140AZ-SPC / UES140BZ-SPC | 15 | 80–264 | 5 | 0.01–3.00 | 0 to +40 | CB,UL,TUV |
| UES140AZ-SPC / UES140BZ-SPC | 27 | 80–264 | 9 | 0.01–3.00 | 0 to +40 | CB,UL,TUV |
| UES140AZ-SPC / UES140BZ-SPC | 36 | 80–264 | 12 | 0.01–3.00 | 0 to +40 | CB,UL,TUV |
| UES140AZ-SPC / UES140BZ-SPC | 45 | 80–264 | 15 | 0.01–3.00 | 0 to +40 | CB,UL,TUV |
| UES140AZ-SPC / UES140BZ-SPC | 100 | 80–264 | 20 | 0.01–5.00 | 0 to +40 | CB,UL,TUV |
UES140AZ-SPC / UES140BZ-SPC | 140 | 80–264 | 28 | 0.01–5.00 | 0 to +40 | CB,UL,TUV |
Across the listed profiles: voltage tolerance ±5%, line regulation ±1%, load regulation ±5%, maximum ripple and noise 300mVp-p, and start-up delay ≤3 seconds.
Mechanical Details

Input Specifications
The 80–264VAC input covers common nominal 100–240VAC mains supplies. Select the AC inlet to match the required installation class, then specify the mains cord for the destination market.
| Input Characteristic | Specification |
| Input voltage range | 80–264VAC |
| Frequency range | 47–63Hz |
| Input current | 2.5A at 90VAC |
| Standby power | ≤0.15W |
| AC Inlet Options |
| C8 / Polarized C8 — Class II; replace Z with 1 |
| C6 — Class I; replace Z with 2 |
| C14 — Class I; replace Z with 3 |
| C18 — Class II; replace Z with 4 |
Environmental & General
| Characteristic | Specification |
| Operating temperature | 0°C to +40°C |
| Storage temperature | −40°C to +80°C |
| Operating humidity | 10–90% RH, non-condensing |
| Storage humidity | 5–90% RH |
| Operating altitude | Up to 5,000m |
| Enclosure rating | IP42 |
| Weight | 340g |
| Energy efficiency | Level VI |
| Average efficiency — 5V | 82.00% |
| Average efficiency — 9V | 86.80% |
| Average efficiency — 12V | 87.50% |
| Average efficiency — 15V | 88.00% |
| Average efficiency — 20V | 90.00% |
| Average efficiency — 28V | 92.00% |
| MTBF | >100,000 hours at 25°C |
Protection
| Protection Function | Specification | Recovery |
| Overload | 105–140% of rated output power | Automatic recovery |
| Short circuit | Trip and restart, hiccup mode | Restart attempts while the fault remains |
Safety Approvals
IEC60601-1
View CertificateEN60601-1
View CertificateEMC
| Item | Medical Specification |
| Conducted emissions | IEC/EN 60601-1-2, CISPR 11 |
| Radiated emissions | IEC/EN 60601-1-2, CISPR 11 |
| Harmonic currents | EN 61000-3-2, Class A |
| Voltage flicker | EN 61000-3-3 |
| Immunity | IEC/EN 60601-1-2 |
| Immunity Test | Standard | Specified Level |
| ESD | IEC 61000-4-2 | ±15kV air, ±8kV contact |
| Radiated immunity | IEC 61000-4-3 | 3V/m |
| EFT/Burst | IEC 61000-4-4 | ±2kV AC port, ±1kV signal ports |
| Surge | IEC 61000-4-5 | ±1kV line-to-neutral; ±1kV line/neutral-to-ground |
| Conducted immunity | IEC 61000-4-6 | 3Vrms, 6Vrms; 0.15–80MHz |
| Magnetic field | IEC 61000-4-8 | 30A/m |
| Dips and interruptions | IEC 61000-4-11 | 0%, 70%, 0% of UT |
Common Applications
These models are options to evaluate for medical equipment designed to accept USB-C PD power. Check the required input profile and total operating demand, including battery charging where applicable, before selecting the adapter.
Frequently Asked Questions
The meter only shows 60W. Should we change the adapter?
Most of the time our device is asleep. Do we really need a large adapter?
Can we order it with a C14 inlet?
We already have 100W cables in stock. Can we use those with AZ?
Can this replace our existing 140W, 19.5V adapter?
Related Models
If this series does not match your equipment’s input requirements, compare the alternatives below before ordering a sample. Consider the required output profile, AC connection and cable arrangement together, particularly when replacing an adapter in an existing equipment package.
Related Resources
Moving to a higher-power PD supply involves more than choosing a larger wattage. These articles provide background on medical USB-C power, patient-protection isolation and the safety documents to review during equipment development.
Medical grade USB-C power supply
An introduction to USB-C power in medical equipment, including connection formats, power delivery and medical-use considerations. Useful background when defining the external power arrangement for a new device.
What does 2 x MOPP mean for a medical power supply?
Both models list 2 MOPP input-to-output isolation. This article explains the term and its role in patient protection, providing context for the isolation requirements in your project specification.
IEC 60601-1 Safety Standards Explained for Power Supply
Review the relationship between equipment classification, insulation and leakage current before comparing approval documents. This overview helps prepare the discussion with your certification engineer and power supply supplier.